Acoustic Materials #
Characterising an absorber or a partition is a laboratory exercise with three distinct instruments behind it. The reverberation room delivers a spectrum of sound-absorption coefficients that ISO 11654 collapses into a single rated number $\alpha_w$ with a letter class. The flow rig — steady or oscillating — measures how hard air is to push through a porous sample, the airflow resistance that governs its low-frequency behaviour (ISO 9053-1/-2). And the impedance tube recovers, at normal incidence, the full complex surface impedance, reflection factor and absorption of a small sample, and — with four microphones — its transmission loss (ISO 10534-1/-2, ASTM E2611). This page covers all three.
1. Sound-absorption rating (ISO 11654) #
ISO 354 measures the sound-absorption coefficient $\alpha_s$ of a material in one-third-octave bands in a reverberation room. ISO 11654:1997 turns that spectrum into a single-number rating comparable across products.
Practical absorption coefficient (Clause 4.1). The one-third-octave data are first grouped into octave bands, each the arithmetic mean of its three thirds:
$$ \alpha_{p,i} = \tfrac{1}{3}\big(\alpha_{i1} + \alpha_{i2} + \alpha_{i3}\big), $$
evaluated to the second decimal and then rounded in steps of $0.05$ (the Clause 4.1 NOTE fixes the rounding, e.g. $0.92 \to 0.90$); rounded means above $1.00$ are set to $1.00$. The five rating bands are 250, 500, 1000, 2000 and 4000 Hz.
Weighted absorption (Clause 4.2). A fixed reference curve ${250{:},0.80,\ 500{:},1.00,\ 1000{:},1.00,\ 2000{:},1.00,\ 4000{:},0.90}$ is shifted downwards, towards the measured $\alpha_p$, in steps of $0.05$ until the sum of the unfavourable deviations — taken only where the measurement lies below the shifted curve, with magnitude $(\text{curve} - \text{measured})$ — is no more than $0.10$. The weighted coefficient $\alpha_w$ is the shifted-curve value read at 500 Hz.
Shape indicators (Clause 4.3). When a practical coefficient exceeds the
shifted curve by $0.25$ or more, a shape indicator is appended: L at 250 Hz,
M at 500 or 1000 Hz, H at 2000 or 4000 Hz (e.g. 0.60(M)).
Absorption class (Table B.1). Finally $\alpha_w$ maps to a class: A (0.90–1.00), B (0.80–0.85), C (0.60–0.75), D (0.30–0.55), E (0.15–0.25), or "not classified" (0.00–0.10). Because $\alpha_w$ is always a multiple of $0.05$ these ranges partition the grid exactly.
The Annex A.2 worked example: the reference curve is shifted down by 0.40 until
the unfavourable deviations sum to 0.05 (≤ 0.10), giving $\alpha_w = 0.60$; the
500 Hz peak overshoots the shifted curve by ≥ 0.25, adding the M indicator, so
the rating is $0.60(\text{M})$, class C.
from phonometry import weighted_absorption, absorption_class
# ISO 11654 Annex A.2 practical coefficients at 250/500/1000/2000/4000 Hz
alpha_p = [0.35, 1.00, 0.65, 0.60, 0.55]
result = weighted_absorption(alpha_p)
print(result.rating_label) # 0.60(M)
print(result.alpha_w) # 0.6
print(result.absorption_class) # C
print(round(result.unfavourable_sum, 2)) # 0.05
# A bare alpha_w also maps straight to its class (Table B.1)
print(absorption_class(0.85)) # B
weighted_absorption accepts the five octave-band $\alpha_p$ values (as a
sequence or a {frequency: value} mapping); pass the fifteen one-third-octave
$\alpha_s$ values to practical_absorption_coefficient first if you are starting
from raw ISO 354 data. The result carries the shifted reference curve and the
per-band deviations, and its .plot() renders the figure above.
2. Airflow resistance (ISO 9053-1/-2) #
The airflow resistance quantifies how strongly a porous material opposes a steady or slowly-oscillating flow. Both parts share the same three quantities and units (ISO 9053-1:2018, Clause 3):
$$ R = \frac{\Delta p}{q_v}\ \left[\text{Pa·s/m}^3\right], \qquad R_s = R,A\ \left[\text{Pa·s/m}\right], \qquad \sigma = \frac{R_s}{d}\ \left[\text{Pa·s/m}^2\right], $$
with $\Delta p$ the pressure difference across the specimen, $q_v$ the volumetric flow, $A$ the cross-section and $d$ the thickness. Note the specific airflow resistance $R_s$ is in Pa·s/m (not Pa·s/m²); the airflow resistivity $\sigma$ is $R_s$ per metre of thickness.
Static method (ISO 9053-1:2018). A steady laminar flow is stepped up and the pressure difference plotted against the linear velocity $u = q_v/A$. A regression of at least second order constrained through the origin, $\Delta p = a,u + b,u^2$, is fitted, and $\Delta p$ and $R_s$ are read at the reference velocity $u = 0.5\ \text{mm/s}$ (Clause 7.5); the highest velocity must not exceed 15 mm/s. Because $R_s = \Delta p/u = a + b,u$, the linear term $a$ is the zero-velocity specific airflow resistance.
The slightly super-linear pressure drop is fitted through the origin; the specific airflow resistance is the fit read at 0.5 mm/s.
import numpy as np
from phonometry import static_airflow_resistance
area = np.pi * 0.05**2 # 100 mm diameter cell [m^2]
u = np.array([0.5, 1, 2, 4, 8, 12]) * 1e-3 # linear velocity [m/s]
dp = 1.6e4 * u + 4.0e5 * u**2 # measured pressure drop [Pa]
r = static_airflow_resistance(u, dp, area=area, thickness=0.05)
print(round(r.specific_resistance)) # 16200 R_s [Pa*s/m]
print(round(r.resistivity)) # 324000 sigma [Pa*s/m^2]
print(round(r.linear_coefficient)) # 16000 a = R_s at u -> 0
Alternating method (ISO 9053-2:2020). A piston oscillating at 1–4 Hz drives an alternating flow into a cavity terminated either by the specimen or by an airtight plug; the resistance follows from the sound-pressure-level difference between the two terminations (Formula (2)):
$$ R = \frac{\kappa',P_S}{2\pi f,V}\cdot\frac{h_t}{h_s}\cdot 10^{(L_{p,s} - L_{p,t})/20}, $$
where $\kappa'$ is the effective ratio of specific heats. Heat conduction between the oscillating air and the cavity walls makes the compression not fully adiabatic; the normative Annex A corrects $\kappa$ down to
$$ \kappa' = \frac{\kappa}{\sqrt{1 + (\kappa-1)\tfrac{S}{V}b + \tfrac{1}{2}\big((\kappa-1)\tfrac{S}{V}b\big)^2}}, \qquad b = \sqrt{\frac{2 c_0 l_h}{\omega}},\quad l_h = \frac{k_a}{\rho_0 c_0 C_P}, $$
with $S$ and $V$ the cavity surface and volume and $b$ the thermal boundary-layer thickness. For the Annex A.3 example cavity this gives $\kappa' = 1.370$, about 2 % below the adiabatic 1.4008.
from phonometry import effective_kappa, alternating_airflow_resistance
# Annex A.3 cavity: closed cylinder 100 mm x 100 mm, piston at 2 Hz
kp = effective_kappa(cavity_surface=0.0471, cavity_volume=7.854e-4, frequency=2.0)
print(round(kp, 3)) # 1.37 effective ratio of specific heats
R = alternating_airflow_resistance(
level_specimen=74.0, level_termination=90.0,
piston_stroke_specimen=14e-3, piston_stroke_termination=1.4e-3,
frequency=2.0, cavity_volume=7.854e-4, kappa_prime=kp,
)
print(round(R)) # airflow resistance R [Pa*s/m^3]
Pass the effective_kappa result to alternating_airflow_resistance for an
Annex-A-conforming figure; its kappa_prime argument otherwise defaults to the
uncorrected adiabatic 1.4. The call warns (via AirflowResistanceWarning) when
the piston frequency leaves 1–4 Hz or the Formula (3)/(4) validity criteria fail.
3. Impedance tube (ISO 10534-1/-2, ASTM E2611) #
The impedance tube measures a small sample at normal incidence. Two standards share the geometry but differ in method and sign convention, so the library keeps their helpers separate and never mixes them.
Standing-wave-ratio method (ISO 10534-1). A probe traverses the standing wave and reads the level difference $\Delta L = L_\text{max} - L_\text{min}$ between a pressure maximum and the adjacent minimum. The standing-wave ratio, reflection magnitude and absorption follow in closed form:
$$ s = 10^{\Delta L/20}, \qquad |r| = \frac{s-1}{s+1}, \qquad \alpha = 1 - |r|^2. $$
A small level difference means a near-perfect absorber; a level difference of 9.54 dB gives $s = 3$, $|r| = 0.5$ and $\alpha = 0.75$.
from phonometry import standing_wave_absorption, standing_wave_ratio_from_level
s = float(standing_wave_ratio_from_level(9.542)) # level difference [dB] -> SWR
print(round(s, 2)) # 3.0
print(round(float(standing_wave_absorption(s)), 2)) # 0.75
Transfer-function method (ISO 10534-2). Two fixed microphones measure the complex transfer function $H_{12}$; from it the reflection factor at the sample face, the absorption and the normalised surface impedance follow (Eqs. (17)–(19)):
$$ r = \frac{H_{12} - H_I}{H_R - H_{12}},e^{,2 j k_0 x_1}, \qquad \alpha = 1 - |r|^2, \qquad \frac{Z}{\rho c_0} = \frac{1+r}{1-r}, $$
with $H_I = e^{-j k_0 s}$, $H_R = e^{+j k_0 s}$, microphone spacing $s$ and $x_1$ the distance from the sample to the farther microphone.
import numpy as np
from phonometry import (
tube_wavenumber, reflection_factor,
absorption_from_reflection, normalized_surface_impedance,
)
f = np.array([500.0, 1000.0, 1800.0])
x1, spacing, c0 = 0.12, 0.03, 343.2
k0 = tube_wavenumber(f, c0)
# A measured transfer function H12 (here synthesised from r = 0.3 - 0.4j)
target = 0.3 - 0.4j
x2 = x1 - spacing
h12 = (np.exp(1j*k0*x2) + target*np.exp(-1j*k0*x2)) / \
(np.exp(1j*k0*x1) + target*np.exp(-1j*k0*x1))
r = reflection_factor(h12, spacing=spacing, x1=x1, wavenumber=k0)
print(np.round(absorption_from_reflection(r), 3)) # [0.75 0.75 0.75]
print(np.round(normalized_surface_impedance(r), 2)) # Z / rho c0
The high-level two_microphone_impedance wraps this chain and returns an
ImpedanceTubeResult with absorption, reflection factor, surface impedance and
normalised impedance, applying the plane-wave frequency-range check and optional
tube attenuation; correct any microphone mismatch beforehand with
apply_mic_calibration.
Transmission loss (ASTM E2611). With four microphones — two upstream, two downstream of the sample — a two-load (or one-load) measurement recovers the sample's transfer matrix, whose entries give the normal-incidence transmission loss, reflection and wavenumber:
$$ \mathrm{TL} = 20\log_{10}\left|\frac{T_{11} + T_{12}/\rho c + \rho c,T_{21} + T_{22}}{2}\right|. $$
import numpy as np
from phonometry import air_layer_transfer_matrix
# An air layer is a known transfer matrix; TL = 0 dB (nothing is lost)
f = np.array([500.0, 1000.0, 2000.0])
k0 = 2*np.pi*f / 343.2
rho_c = 1.186 * 343.2
tm = air_layer_transfer_matrix(thickness=0.05, wavenumber=k0,
characteristic_impedance=rho_c)
print(np.round(tm.transmission_loss(rho_c), 6)) # [0. 0. 0.]
transfer_matrix_two_load / transfer_matrix_one_load build the
TransferMatrix from measured four-microphone wave amplitudes; its methods
(transmission_loss, reflection_hard_backed, absorption_hard_backed,
characteristic_impedance_material, material_wavenumber) then read off the
ASTM E2611 quantities.
Standards: ISO 11654:1997, ISO 9053-1:2018, ISO 9053-2:2020, ISO 10534-1:1996, ISO 10534-2:1998 and ASTM E2611-19. Every equation is derived from the standard text; the conformance report validates the library against the standards' own worked examples (ISO 11654 Annex A, ISO 9053-2 Annex A.3) and closed-form identities.