Sound Intensity (p-p method) #

Sound pressure tells you how loud a point is; sound intensity tells you where the energy is going. It is the acoustic power flux (W/m²), a signed vector quantity — which is why intensity probes can localize sources, separate them from background noise and measure sound power in situ (ISO 9614) where a pressure measurement alone cannot.

The two-microphone principle (IEC 61043) #

A p-p probe holds two matched microphones a small distance Δr apart. The pressure at the probe center is their mean, and the particle velocity comes from the pressure gradient (Euler's equation, finite-difference form):

$$ p = \frac{p_1 + p_2}{2}, \qquad u = -\frac{1}{\rho_0\ \Delta r}\int (p_2 - p_1)\ dt, \qquad I = \overline{p\ u} $$

In practice the estimator works in the frequency domain through the cross-spectrum of the two channels (the standard's equivalent form):

$$ I(f) = -\ \frac{\mathrm{Im}\lbrace G_{12}(f)\rbrace}{2\pi f\ \rho_0\ \Delta r} $$

Two-microphone p-p intensity probe with the spacer distance and the measurement axis

import numpy as np
from phonometry import sound_intensity

fs = 48000
rng = np.random.default_rng(0)
# The two probe-microphone pressures in Pa, p1 closest to the source.
#   In a real measurement these are your two calibrated probe recordings;
#   synthesized here (p2 = p1 delayed one sample) so the guide runs.
p1 = 0.02 * rng.standard_normal(fs)
p2 = np.concatenate(([0.0], p1[:-1]))   # p2 = p1 delayed one sample

res = sound_intensity(p1, p2, fs, spacing=0.012, fraction=3,
                      limits=[100, 2500])
print(res.total_intensity_level, res.total_direction)      # LI [dB], ±1
print(res.frequency, res.intensity_level)                  # per band
res.plot()   # Lp vs LI per band + the pressure-intensity index (needs matplotlib)

Third-octave pressure and intensity levels for a plane progressive wave versus a standing wave

Left: in a plane progressive wave all pressure is transported — L_I ≈ L_p. Right: a standing wave carries (almost) no net energy — the pressure is high but the intensity collapses. The gap L_p − L_I is the pressure-intensity index, the fundamental quality indicator of every intensity measurement.

Show the code for this figure
import matplotlib.pyplot as plt

# res is the IntensityResult computed in the example above.
# One line — Lp vs LI per band with the pressure-intensity index on a twin axis:
res.plot()
plt.show()

# By hand, from the per-band fields the result carries — mirroring what
# IntensityResult.plot() draws (bar label, merged twin-axis legend, δpI title):
fig, ax = plt.subplots()
ax.semilogx(res.frequency, res.pressure_level, "o-", label="Pressure level Lp")
ax.semilogx(res.frequency, res.intensity_level, "s--", label="Intensity level LI")
ax.set_xlabel("Frequency [Hz]")
ax.set_ylabel("Level [dB]")
twin = ax.twinx()
twin.bar(res.frequency, res.pressure_intensity_index,
         width=res.frequency * 0.2, color="#2ca02c", alpha=0.25,
         label="δpI = Lp − LI")
twin.set_ylabel("Pressure-intensity index δpI [dB]")
# Merge both axes' handles into a single legend, exactly as .plot() does:
lines, labels = ax.get_legend_handles_labels()
tlines, tlabels = twin.get_legend_handles_labels()
ax.legend(lines + tlines, labels + tlabels)
ax.set_title(f"Lp vs LI  (total δpI = {res.total_pressure_intensity_index:.1f} dB)")
plt.show()

Knowing when to trust the number #

Two physical limits bound every p-p measurement, and the result object carries both:

  • High frequency: the finite-difference gradient underestimates I by $\sin(k\Delta r)/(k\Delta r)$ — verified in CI against IEC 61043 Table 3. IntensityResult.bias_correction provides the factor and max_valid_frequency (≈ 0.1·c/Δr; 2.9 kHz for a 12 mm spacer) the practical ceiling. Larger spacers reach lower frequencies, smaller ones higher.
  • Reactive fields: when pressure_intensity_index (F2 in ISO 9614-1) approaches the probe's residual index δ_pI0, phase errors dominate. The ISO 9614-1 Annex A field indicators and the dynamic-capability criterion are available directly:
from phonometry import field_indicators, dynamic_capability_index

# Per-position measurements over the ISO 9614-1 measurement surface
pressure_levels = np.array([74.1, 73.8, 74.5, 73.2])       # Lp per position (dB)
normal_intensity = np.array([1.2e-5, 1.0e-5, 1.4e-5, 0.9e-5])  # signed In per position (W/m²)

fi = field_indicators(pressure_levels, normal_intensity)   # F2, F3, F4
ld = dynamic_capability_index(18.0)   # δpI0 = 18 dB → Ld = δpI0 − K
ok = ld > fi.f2                                            # criterion 1

sound_intensity() parameters #

Parameter Type Units Range / default Notes
p1, p2 1D arrays Pa equal length Microphone closer to the source first; reversing them flips the sign
fs int Hz > 0
spacing float m > 0 Microphone separation Δr (typ. 6/12/50 mm)
rho float kg/m³ default 1.204 Air density
c float m/s default 343.0 Speed of sound (bias/validity estimates)
fraction int, optional — 1, 3 or None (default) Octave/third-octave band integration
limits list, optional Hz default library band range Band analysis limits
bias_correct bool — default False Apply the per-bin $(k\Delta r)/\sin(k\Delta r)$ correction (IEC 61043 §7.3) before summing, so band/broadband totals stop under-reading as $f \to$ max_valid_frequency; bins past the first null are left uncorrected. The per-band bias_correction factor is reported either way

See Theory for the derivations and Calibration for absolute scaling of the two channels.