# Sound Intensity (p-p method)
Sound *pressure* tells you how loud a point is; sound **intensity** tells
you where the energy is *going*. It is the acoustic power flux (W/m²), a
signed vector quantity — which is why intensity probes can localize sources,
separate them from background noise and measure sound power in situ
(ISO 9614) where a pressure measurement alone cannot.
## The two-microphone principle (IEC 61043)
A p-p probe holds two matched microphones a small distance Δr apart. The
pressure at the probe center is their mean, and the particle velocity comes
from the pressure *gradient* (Euler's equation, finite-difference form):
$$
p = \frac{p_1 + p_2}{2}, \qquad
u = -\frac{1}{\rho_0\ \Delta r}\int (p_2 - p_1)\ dt, \qquad
I = \overline{p\ u}
$$
In practice the estimator works in the frequency domain through the
cross-spectrum of the two channels (the standard's equivalent form):
$$
I(f) = -\ \frac{\mathrm{Im}\lbrace G_{12}(f)\rbrace}{2\pi f\ \rho_0\ \Delta r}
$$
```python
import numpy as np
from phonometry import sound_intensity
fs = 48000
rng = np.random.default_rng(0)
# The two probe-microphone pressures in Pa, p1 closest to the source.
# In a real measurement these are your two calibrated probe recordings;
# synthesized here (p2 = p1 delayed one sample) so the guide runs.
p1 = 0.02 * rng.standard_normal(fs)
p2 = np.concatenate(([0.0], p1[:-1])) # p2 = p1 delayed one sample
res = sound_intensity(p1, p2, fs, spacing=0.012, fraction=3,
limits=[100, 2500])
print(res.total_intensity_level, res.total_direction) # LI [dB], ±1
print(res.frequency, res.intensity_level) # per band
res.plot() # Lp vs LI per band + the pressure-intensity index (needs matplotlib)
```
*Left: in a plane progressive wave all pressure is transported —
L_I ≈ L_p. Right: a standing wave carries (almost) no net energy — the
pressure is high but the intensity collapses. The gap L_p − L_I is the
**pressure-intensity index**, the fundamental quality indicator of every
intensity measurement.*
Show the code for this figure
```python
import matplotlib.pyplot as plt
# res is the IntensityResult computed in the example above.
# One line — Lp vs LI per band with the pressure-intensity index on a twin axis:
res.plot()
plt.show()
# By hand, from the per-band fields the result carries — mirroring what
# IntensityResult.plot() draws (bar label, merged twin-axis legend, δpI title):
fig, ax = plt.subplots()
ax.semilogx(res.frequency, res.pressure_level, "o-", label="Pressure level Lp")
ax.semilogx(res.frequency, res.intensity_level, "s--", label="Intensity level LI")
ax.set_xlabel("Frequency [Hz]")
ax.set_ylabel("Level [dB]")
twin = ax.twinx()
twin.bar(res.frequency, res.pressure_intensity_index,
width=res.frequency * 0.2, color="#2ca02c", alpha=0.25,
label="δpI = Lp − LI")
twin.set_ylabel("Pressure-intensity index δpI [dB]")
# Merge both axes' handles into a single legend, exactly as .plot() does:
lines, labels = ax.get_legend_handles_labels()
tlines, tlabels = twin.get_legend_handles_labels()
ax.legend(lines + tlines, labels + tlabels)
ax.set_title(f"Lp vs LI (total δpI = {res.total_pressure_intensity_index:.1f} dB)")
plt.show()
```
## Knowing when to trust the number
Two physical limits bound every p-p measurement, and the result object
carries both:
- **High frequency**: the finite-difference gradient underestimates I by
$\sin(k\Delta r)/(k\Delta r)$ — verified in CI against IEC 61043 Table 3.
`IntensityResult.bias_correction` provides the factor and
`max_valid_frequency` (≈ 0.1·c/Δr; 2.9 kHz for a 12 mm spacer) the
practical ceiling. Larger spacers reach lower frequencies, smaller ones
higher.
- **Reactive fields**: when `pressure_intensity_index` (F2 in ISO 9614-1)
approaches the probe's residual index δ_pI0, phase errors dominate. The
ISO 9614-1 Annex A field indicators and the dynamic-capability criterion
are available directly:
```python
from phonometry import field_indicators, dynamic_capability_index
# Per-position measurements over the ISO 9614-1 measurement surface
pressure_levels = np.array([74.1, 73.8, 74.5, 73.2]) # Lp per position (dB)
normal_intensity = np.array([1.2e-5, 1.0e-5, 1.4e-5, 0.9e-5]) # signed In per position (W/m²)
fi = field_indicators(pressure_levels, normal_intensity) # F2, F3, F4
ld = dynamic_capability_index(18.0) # δpI0 = 18 dB → Ld = δpI0 − K
ok = ld > fi.f2 # criterion 1
```
### `sound_intensity()` parameters
| Parameter | Type | Units | Range / default | Notes |
| :--- | :--- | :--- | :--- | :--- |
| `p1`, `p2` | 1D arrays | Pa | equal length | Microphone closer to the source first; reversing them flips the sign |
| `fs` | int | Hz | > 0 | |
| `spacing` | float | m | > 0 | Microphone separation Δr (typ. 6/12/50 mm) |
| `rho` | float | kg/m³ | default `1.204` | Air density |
| `c` | float | m/s | default `343.0` | Speed of sound (bias/validity estimates) |
| `fraction` | int, optional | — | `1`, `3` or `None` (default) | Octave/third-octave band integration |
| `limits` | list, optional | Hz | default library band range | Band analysis limits |
| `bias_correct` | bool | — | default `False` | Apply the per-bin $(k\Delta r)/\sin(k\Delta r)$ correction (IEC 61043 §7.3) before summing, so band/broadband totals stop under-reading as $f \to$ `max_valid_frequency`; bins past the first null are left uncorrected. The per-band `bias_correction` factor is reported either way |
See [Theory](theory.md) for the derivations and [Calibration](calibration.md)
for absolute scaling of the two channels.