Calibration and dBFS #

phonometry can return results in physical Sound Pressure Level (dB SPL) or digital decibels relative to Full Scale (dBFS).

Why calibrate? The theory #

A digital recording only knows numbers: a full-scale sine wave is ±1.0 regardless of whether it was a whisper or a jet engine. To report physical sound pressure levels the chain microphone → preamplifier → ADC must be characterized by a single number, the sensitivity factor S, that converts digital units into pascals:

$$ p(t) = S , x(t) \qquad S = \frac{p_\text{ref} \cdot 10^{L_\text{cal}/20}}{\tilde{x}_\text{ref}} $$

where $L_\text{cal}$ is the calibrator's level (typically 94 dB, i.e. 1 Pa), $p_\text{ref} = 20\ \mu\text{Pa}$ and $\tilde{x}_\text{ref}$ is the RMS of the recorded calibration tone in digital units. calculate_sensitivity() is exactly that equation. The factor is valid as long as nothing in the chain changes — touch the gain knob and you must recalibrate.

Calibration chain: sound calibrator coupled on the microphone, preamplifier, ADC and calculate_sensitivity producing pascals per digital unit

Physical Calibration (Sound Level Meter) #

Calibrator tone94 dB @ 1 kHz(IEC 60942)Recordingref_signalcalculate_sensitivity()calibration_factor(digital units → Pa)octavefilter / leq / laeq /ln_levelsMeasurementrecordingLevels in dB SPL(re 20 µPa)

To get accurate SPL measurements from a digital recording, you must first calculate the sensitivity of your measurement chain using a reference tone (e.g., 94 dB @ 1 kHz).

from phonometry import octavefilter, calculate_sensitivity

# 1. Record your 94dB calibrator signal
# ref_signal = ... (your recording)

# 2. Calculate sensitivity factor
sensitivity = calculate_sensitivity(ref_signal, target_spl=94.0)

# 3. Apply calibration to your measurements
spl, freq = octavefilter(signal, fs, calibration_factor=sensitivity)
# Now 'spl' values are in real-world dB SPL!

The same calibration_factor works across the whole library: octavefilter, OctaveFilterBank, leq, laeq and ln_levels.

Calibrator assumptions (IEC 60942) #

calculate_sensitivity assumes the reference recording comes from an acoustic calibrator as specified by IEC 60942 (classes LS, 1 and 2):

  • The default target_spl=94.0 matches the common 94 dB @ 1 kHz calibrator output (the standard requires the principal level to be at least 90 dB re 20 µPa; 94 dB and 114 dB are the usual choices).
  • The resulting sensitivity inherits the calibrator's class tolerance — e.g. ±0.4 dB for a class 1 calibrator between 160 Hz and 1.25 kHz (IEC 60942 Table 1) — plus the RMS estimation error of your recording.
  • IEC 60942 specifies the generated level as a 20 s average: record a few seconds of stable tone (excluding handling noise at the start/end) for the RMS estimate to converge.

Automatic stability validation #

When you pass the sample rate (and validate=True, the default), calculate_sensitivity(ref, fs=fs) checks the recording the way IEC 60942:2017 checks the calibrator itself (5.3.3): the short-term level fluctuation — the absolute difference between each of the maximum and minimum F-time-weighted levels and the mean level — must not exceed the Table 2 class 1 limit for the calibrator's nominal frequency (0.07 dB at and above 160 Hz, relaxed to 0.10 dB below 160 Hz and 0.20 dB at or below 63 Hz, where the F time-weighting itself ripples). Pass frequency= to select the right row for non-1 kHz calibrators. A CalibrationWarning flags badly coupled microphones or handling noise before they silently corrupt every calibrated level. The recording must be at least 2 s long (1 s for the F-integrator to settle plus 1 s of settled envelope); shorter recordings get a warning instead of an unreliable verdict. Without fs the check is skipped. Override the limit with max_fluctuation_db or disable with validate=False.

The check catches exactly what ruins field calibrations — a loose coupler, wind, handling noise:

F-weighted level of a stable calibration tone versus a 3 percent amplitude-modulated one against the plus-minus 0.07 dB IEC 60942 class 1 limit

calculate_sensitivity() parameters #

Parameter Type / shape Units Range / default Notes
ref_signal 1D/2D array digital units non-empty, non-silent Recording of the calibration tone only (trim handling noise)
target_spl float dB re 20 µPa default 94.0 The calibrator's nominal level (114 dB calibrators: pass 114.0)
ref_pressure float Pa default 2e-5 Reference pressure p₀; rarely changed
fs int, optional Hz > 0; default None Required for the stability validation; omit to skip it
validate bool — default True Emit CalibrationWarning on unstable/short recordings
max_fluctuation_db float, optional dB default None → Table 2 class 1 Explicit override of the stability limit
frequency float Hz default 1000.0 Calibrator's nominal frequency; selects the IEC 60942 Table 2 row

Returns the sensitivity factor (float) to pass as calibration_factor= to octavefilter, leq, laeq, ln_levels, lc_peak, sel and the dose functions.

Digital Analysis (dBFS) #

If you are working with digital audio files (e.g., WAV, FLAC) and want to analyze levels relative to Full Scale rather than physical pressure, you can use the dbfs=True parameter.

In this mode:

  • 0 dBFS corresponds to a numeric signal level of 1.0 (RMS or Peak).
  • calibration_factor does not apply (dBFS is relative to digital full scale).
  • Useful for analyzing headroom, digital mastering, or normalized signals.
# Assume 'signal' is normalized between -1.0 and 1.0
spl_dbfs, freq = octavefilter(signal, fs, dbfs=True)
# Results will be negative (e.g., -20 dBFS)

RMS vs Peak Levels #

phonometry supports two measurement modes to align with professional software like BK:

  • RMS (mode='rms'): Energy-based level (standard).
  • Peak (mode='peak'): Absolute maximum value reached in the frame (Peak-holding).
# Measure peak-holding levels for impact analysis
spl_peak, freq = octavefilter(signal, fs, mode='peak')

Note

mode='peak' measures the absolute maximum of the filtered band signal, which includes the filter's onset transient (overshoot). Signals that start abruptly may read up to ~1 dB high. This is inherent to IIR band filters (an analog SLM behaves the same way), not a processing artifact.

Integer audio input #

Integer signals (e.g. int16 from scipy.io.wavfile.read) are converted to float64 internally before any squaring, so calibration and level results are identical whether you pass the raw integer array or a float conversion.