← [Documentation index](README.md)
# Calibration and dBFS
phonometry can return results in physical **Sound Pressure Level (dB SPL)** or
digital **decibels relative to Full Scale (dBFS)**.
## Why calibrate? The theory
A digital recording only knows *numbers*: a full-scale sine wave is ±1.0
regardless of whether it was a whisper or a jet engine. To report physical
sound pressure levels the chain microphone → preamplifier → ADC must be
characterized by a single number, the **sensitivity factor** S, that converts
digital units into pascals:
$$
p(t) = S \ x(t) \qquad
S = \frac{p_\text{ref} \cdot 10^{L_\text{cal}/20}}{\tilde{x}_\text{ref}}
$$
where $L_\text{cal}$ is the calibrator's level (typically 94 dB, i.e. 1 Pa),
$p_\text{ref} = 20\ \mu\text{Pa}$ and $\tilde{x}_\text{ref}$ is the RMS of
the recorded calibration tone in digital units. `calculate_sensitivity()` is
exactly that equation. The factor is valid as long as nothing in the chain
changes — touch the gain knob and you must recalibrate.
<picture><source media="(prefers-color-scheme: dark)" srcset="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_calibration_setup_dark.svg"><img src="https://raw.githubusercontent.com/jmrplens/phonometry/main/.github/images/diagram_calibration_setup.svg" alt="Calibration chain: sound calibrator coupled on the microphone, preamplifier, ADC and calculate_sensitivity producing pascals per digital unit" width="92%"></picture>
## Physical Calibration (Sound Level Meter)
```mermaid
flowchart LR
A["Calibrator tone\n94 dB @ 1 kHz\n(IEC 60942)"] --> B["Recording\nref_signal"]
B --> C["calculate_sensitivity()"]
C --> D["calibration_factor\n(digital units → Pa)"]
D --> E["octavefilter / leq / laeq / ln_levels"]
F["Measurement\nrecording"] --> E
E --> G["Levels in dB SPL\n(re 20 µPa)"]
```
To get accurate SPL measurements from a digital recording, you must first
calculate the sensitivity of your measurement chain using a reference tone
(e.g., 94 dB @ 1 kHz).
```python
from…